An AED spots shockable rhythms mid-compression without new sensors: six noise conditions pass the adjudicable AHA criteria at 0 dB artifact
Public defibrillators must pause chest compressions to read the ECG because compressions bury it in artifact; a three-network cascade classifies the contaminated signal directly using only the ECG channel the device already has, with a network biased toward the proposed outcome verifying each decision and disagreement yielding an indeterminate result. On an independent manufacturer-provided cohort of 464 subjects, at 0 dB compression artifact with an eight-second window across six noise conditions, sinus-rhythm specificity was 99.20–99.43% and specificity for other non-shockable rhythms 95.48–96.20%, with only 1.07–1.48% of segments indeterminate; at the harder artifact level the two adjudicable criteria are no longer met.